2007005/10/001 Labom United States Sales
If you are looking for a reliable supplier where you can purchase the 2007005/10/001 PRESSURE TRANSMITTER PASCAL CV, you are in the right place. ONRION sells all product groups belonging to the Labom brand in the United States market, with fast delivery times and competitive prices.
Unit price
USD 2,130.00
excl. VAT
Estimated delivery time:
6-8 weeks (Exact delivery time: Determined at the time of quotation)
The price shown above is for 1 unit. The price may vary up or down if you order more units.
VISA
Mastercard
Bank Transfer
Technical Specification
Order Code: CV3110-A1059-F80-H1100-M1-N51-T6-T31-K1085-W4020-W1201-W2
602 DL8080-HY-D58-G7-A400-L22-W1020
CV3110 PRESSURE TRANSMITTER PASCAL CV, for diaphragm seal operation,
hygienic
acc. to data sheet D4-017/22
A1 pressure sensor: piezoresistive for gauge pressure
059 nominal range: 0...16 bar, overload limit 60 bar
measuring range setting: turndown 5:1
accuracy: <= 0.15% of measuring span (linearity, hysteresis and
repeatability)
F80 measuring range: setting see below
lower range value: - 1,000
upper range value: + 15,000
unit: bar
H11 output signal: 4...20 mA, rising characteristic (standard)
0 setting damping: 0.0 sec. (standard)
0 setting alarm state: < 3.6 mA (standard)
M1 display module: without
screw cap closed, without window
N5 switching module: with 2 floating channels, pluggable, electrically
isolated at all sides, switching capacity 30 V DC / 0.5 A
1 switching module setting: at the factory, see below
1. switch contact:
2. switch contact:
T6 case: hygienic design with screw cap
material: stainless steel
degree of protection: IP 66 incl, vent filter, stainless steel
T31 electrical connection: circular connector M 12x1 (8 pin)
K1085 design standard
W4020 ADDITIONAL DESIGN: marking (adhesive label), see below
W1201 INSPECTION CERTIFICATE DIN EN 10204 3.1
calibration certificate with 5 measuring points
W2602 functional safety per EN 61508, classification in SIL2
TÜV-Reg.-no. 44 799 13190201
DL8080 DIAPHRAGM SEAL, VARIVENT® for VARINLINE® access unit
see data sheet D5-024-1
HY hygienic design per EHEDG Doc.8 and ASME BPE SF3:
surface roughness (wetted parts)
- turned parts Ra <= 0.76 µm
- diaphragm Ra <= 0.38 µm
- weld seam Ra <= 0.76 µm
D58 VARIVENT® connection Form N (D = 68) for VARINLINE® access unit, PN 25
G7 diaphragm material: stainless steel mat.-no. 1.4435 (316L)
A400 connection of meas. instrument directly welded
L22 system filling: Synthetic oil, free of silicon FD1
acc. to FDA 21 CFR 178.3570 / H1-registered
operating temperature range -10...+140 °C
W1020 Components marked acc. to material inspection certificate
per EN 10204-3.1, wetted parts
Spare part for fabrication no. 2007005/10/001