MS-CG-5000 Msa United States Sales
If you are looking for a reliable supplier where you can purchase the MS-CG-5000 industrial product, you are in the right place. ONRION sells all product groups belonging to the Msa brand in the United States market, with fast delivery times and competitive prices.
Technical Specification
CHILLGARD 5000
The Chillgard 5000 is a high-resolution
Refrigerant analyzer with unrivalled accuracy
of 1 ppm and excellent long-term stability
- Real-time monitoring up to 1 ppm with patented
photoacoustic infrared technology (PAIR).
- Hassle-free operation with future-oriented
sensor diagnostics and predictive maintenance.
- pump module with 4, 8 or 16 suction points
- Suction length up to 121 m optimal, max 381 m
- Integrated refrigerant library on site with up to
to six selectable refrigerants to be monitored
- Clarifying agent gases:
R11, R12, R22, R123, R134A, R401A, R404A, R407A,
R407C, R407F, R410A, R422A, R422D, R427A, R507,
R1233zd (E), R1234yf, R1234ze, R-513A, R-514A
- Detection limit 1 ppm for all listed gases
- Measurement principle: infrared photoacoustic
- Measurement range: 0 - 1000 ppm
- Adjustable alarms per measurement channel
- 5 relays 8A 250 VAC/24 VDC for faults, pre-warning,
Warning level, alert level, and external horn
- Display: 7 inch touch display for easy operation
and full text display can be selected in six languages
- Output: 4-20 mA and 0-10 V
Acoustic alarm: max. 95+-5 dB at 61 cm
- Power consumption: 100-240 VAC/200W
- Dimensions: H x W x D 396 x 345 x 193 mm
HS Code: 90271010 | Country of Origin: USA
Variant: A-CG5000-1-8M-0-0-M
(MSA 10217009, metric version)